News Release

Sepsis outbreak at L.A. County dialysis center prompts public health investigation

Improper disinfection of reusable medical device identified as cause

Peer-Reviewed Publication

Association for Professionals in Infection Control

San Antonio, Texas, June 5, 2012 – Three patients with chronic kidney failure treated at a dialysis center in Los Angeles County, California contracted a bacterial infection in the blood (sepsis) caused by improper cleaning and disinfection of a reusable medical device called a dialyzer – an artificial kidney.

Described in a poster presented at the 39th Annual Educational Conference and International Meeting of the Association for Professionals in Infection Control and Epidemiology (APIC), the County of Los Angeles Department of Public Health, conducted an investigation led by public health nurse L'Tanya English, RN, MPH, who found that the bacteria infecting the three patients were genetically linked. These patients were infected with Stenotrophomonas maltophilia (S. maltophilia), a rare type of gram-negative bacteria.

Two of these patients were also positive for Candida parapsilosis (C. parapsilosis), a fungus that can cause sepsis in immune-compromised patients. One of these patients was positive for C. parapsilosis in the dialyzer only, and one patient was positive for Candida in the blood and in the dialyzer, which was genetically traced back to the same fungus in a faucet in the reprocessing room, where the dialyzers are disinfected and sanitized. The infections were reported to the health department in August 2011.

Two patients developed fevers and were hospitalized. One patient was assessed and treated as an outpatient; all patients later recovered.

The County of Los Angeles Department of Public Health became aware of the situation when a hospital in southern California reported an outbreak of sepsis tied to one dialysis center. During the course of their investigation, they discovered that all of the cases used the same type of dialyzer with a removable component – an O-ring header. These three patients were the only ones in the facility to use this type of dialyzer. In response to this outbreak, the facility decided to discontinue use of multi-use dialyzers with O-ring headers.

"Hemodialysis technology is life-saving, but carries a high risk of infection, regardless of the type of dialyzer used," said English. "Dialysis centers must work to reduce the risk of infection for their patients by ensuring proper cleaning and disinfection procedures are being followed throughout the facility. If multi-use dialyzers with removable headers and O-rings are used, processes to ensure proper disinfection must be in place."

The County of Los Angeles Department of Public Health is working with state and federal partners to conduct outreach to dialysis centers to decrease dialysis-associated infections and will discuss lessons learned from the investigation at the APIC Annual Conference.

"Contaminated O-rings have been previously implicated in dialysis-associated infection outbreaks. This report underscores the need for adequate infection prevention training in dialysis settings, as well as the critical partnership between public health departments and infection preventionists in hospitals and outpatient settings," said Michelle Farber, RN, CIC, APIC 2012 president. "Collaboration with public health is essential to pinpoint the cause of infection outbreaks and improve infection prevention practices across all healthcare settings."

Hemodialysis is a life-saving procedure that uses an artificial kidney, or dialyzer, to remove waste from the blood when the kidneys no longer work. It is most often the treatment for end-stage renal disease. Following cardiovascular disease, infection is the second highest cause of death for hemodialysis patients.

The most recent draft of the U.S. Department of Health and Human Services' National Action Plan to Prevent Healthcare-associated Infections: Roadmap to Elimination includes a revised chapter on efforts to prevent and reduce healthcare-associated infections in end-stage renal disease patients.

In an effort to establish best practices for protecting patients undergoing hemodialysis, APIC published a Guide to the Elimination of Infections in Hemodialysis and has an archived webinar on dialysis event surveillance and reporting.

###

The APIC 2012 Annual Conference, June 4-6 in San Antonio, is the most comprehensive infection prevention conference in the world, featuring more than 100 educational sessions and workshops led by top health experts. The conference is designed to help attendees translate the latest science into practical strategies for the reduction of healthcare-associated infections. The Twitter hashtag #APIC2012 is being used for the meeting.

Poster Presentation #9-136 – Outbreak investigation at a dialysis center associated with a multi-use dialyzer with removable headers and O-rings, Los Angeles County.

About APIC

APIC's mission is to create a safer world through prevention of infection. The association's more than 14,000 members direct infection prevention programs that save lives and improve the bottom line for hospitals and other healthcare facilities. APIC advances its mission through patient safety, implementation science, competencies and certification, advocacy, and data standardization. Follow APIC on Twitter: http://twitter.com/apic.


Disclaimer: AAAS and EurekAlert! are not responsible for the accuracy of news releases posted to EurekAlert! by contributing institutions or for the use of any information through the EurekAlert system.